A noninvasive estimation of central blood-pressure is required by the researchers or the medical practitioners in the cardiovascular treatment. In this study, first of all, a cardiovascular simulator based on the electrical circuits is described based on the model based approach and then, the representation of aorta model as a pressure transfer function is shown. This representation accelerates the simulator and then, makes it possible to estimate the central blood-pressure from a wearable gouge measurement in a real-time way. The system consisting of a main station, gateways and wearable blood-pressure gauges using one-board microcomputer is actually implemented. The functions and performances of the system are experimentally evaluated and the obtained results show the feasibility of central blood-pressure estimation and the localization of some examinees having wearable gauges with the practical accuracy and speed.
Published in | Journal of Health and Environmental Research (Volume 3, Issue 4) |
DOI | 10.11648/j.jher.20170304.12 |
Page(s) | 62-71 |
Creative Commons |
This is an Open Access article, distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution and reproduction in any medium or format, provided the original work is properly cited. |
Copyright |
Copyright © The Author(s), 2017. Published by Science Publishing Group |
Aorta Model, Cardiovascular Simulator, Central Blood-Pressure, IoT, Microprocessor, Transfer Function, Wearable Sensor
[1] | Vital Statistics in JAPAN -The latest trends- Website, http://www.mhlw.go.jp/english/database/db-hw/dl/81-1a2en.pdf. |
[2] | Automatic blood pressure monitor / arm / with central aortic pressure monitor BP+, Website http://www.medicalexpo.com/prod/uscom/product-70427-543975.html. |
[3] | Biosignalplux, wearable body sensing platform, Website, http://biosignalsplux.com/index.php/en/. |
[4] | S. Takahashi, K. Sakawa, Y. Shiraishi and H. Miyashita, “Modeling, Simulation and Parameter Estimation of the Cardiovascular System by using Model Based Approach,” SICE 2013 Annual Conference, pp.493-500, Hokkaido, September, 2013. |
[5] | K. Nakajo, Y. Komori, S. Takahashi, K. Motegi, Y. Shiraishi, and H. Miyashita, “Pressure Transfer Function for Aorta Model in Cardiovascular Simulator, Feasibility Study of Wearable Central Blood-Pressure Gauge,” International Conference on Medical Engineering, Health Informatics and Technology (MediTec 2016), Dhaka, December 17-18, 2016. |
[6] | H. Miyashita, M. Sugimachi, T. Sato, T. Kawada, T. Shishido, T. Nakahara, R. Yoshimura, H. Takaki, H. Miyao, and K. Sunagawa, “A novel servo-control system that imposes desired aortic input impedance on situ rat heart,” Am. J. Physiol. Heart Circ. Physiol. 278: H998-H1007, 2000. |
[7] | K. B. Campbell, R. Burattini, D. L. Bell, R. D. Kirkpatrick and G. G. Knowlen, “Time-domain formulation of asymmetric T-tube model of arterial system,” Am. J. Physiol. Heart Circ. Physiol. 258:H1761-H1774, 1990. |
[8] | W. W. Nichols, M. F. O'Rourke, C. Vlachopoulos, “McDonald's Blood Flow in Arteries,” Theoretical, Experimental and Clinical Principles, 6th ed. London: Hodders Arnold, 2011. |
[9] | Math Works, Website, http://www.mathworks.co.jp/products/pfo/. |
[10] | N. Bulusu, J. Heidemann, and D. Estrin, "GPS-less low cost outdoor localization for very small devices," IEEE Personal Communications Magazine, Vo.7, No.5, pp.28–34, October 2000. |
[11] | J. Blumenthal, R. Grossmann, F. Golatowski, and D. Timmermann, "Weighted Centroid Localization in Zigbee-based Sensor Networks," IEEE International Symposium on Intelligent Signal Processing, October, 2007. |
[12] | “RASPBERRY PI3 MODEL B,” https://www.raspberrypi.org/products/raspberry-pi-3-model-b/. |
[13] | “BLE Nano,” http://redbearlab.com/blenano/. |
[14] | J. P. Murgo, N. Westerhof, J. P. Giolma and S. A. Altobelli, “Aortic input impedance in normal man: relationship to pressure waveforms,” Circulation 62: pp.105-116, 1980. |
[15] | N. Guan, Z. G. Wen and K. G. Sun, “Research on Indoor Positioning Algorithm Based on Trilateral Positioning and Taylor Series Expansion,” Proceedings of 2016 International Conference on Computational Modeling, Simulation and Applied Mathematics (CMSAM 2016), 3266, Bangkok, July 2016. |
APA Style
Shuji Takahashi, Yudai Komori, Kyoji Nakajo, Kazuhiro Motegi, Yoichi Shiraishi, et al. (2017). A Feasibility Study of Wearable Central Blood-Pressure Measurement System. Journal of Health and Environmental Research, 3(4), 62-71. https://doi.org/10.11648/j.jher.20170304.12
ACS Style
Shuji Takahashi; Yudai Komori; Kyoji Nakajo; Kazuhiro Motegi; Yoichi Shiraishi, et al. A Feasibility Study of Wearable Central Blood-Pressure Measurement System. J. Health Environ. Res. 2017, 3(4), 62-71. doi: 10.11648/j.jher.20170304.12
AMA Style
Shuji Takahashi, Yudai Komori, Kyoji Nakajo, Kazuhiro Motegi, Yoichi Shiraishi, et al. A Feasibility Study of Wearable Central Blood-Pressure Measurement System. J Health Environ Res. 2017;3(4):62-71. doi: 10.11648/j.jher.20170304.12
@article{10.11648/j.jher.20170304.12, author = {Shuji Takahashi and Yudai Komori and Kyoji Nakajo and Kazuhiro Motegi and Yoichi Shiraishi and Hiroshi Miyashita}, title = {A Feasibility Study of Wearable Central Blood-Pressure Measurement System}, journal = {Journal of Health and Environmental Research}, volume = {3}, number = {4}, pages = {62-71}, doi = {10.11648/j.jher.20170304.12}, url = {https://doi.org/10.11648/j.jher.20170304.12}, eprint = {https://article.sciencepublishinggroup.com/pdf/10.11648.j.jher.20170304.12}, abstract = {A noninvasive estimation of central blood-pressure is required by the researchers or the medical practitioners in the cardiovascular treatment. In this study, first of all, a cardiovascular simulator based on the electrical circuits is described based on the model based approach and then, the representation of aorta model as a pressure transfer function is shown. This representation accelerates the simulator and then, makes it possible to estimate the central blood-pressure from a wearable gouge measurement in a real-time way. The system consisting of a main station, gateways and wearable blood-pressure gauges using one-board microcomputer is actually implemented. The functions and performances of the system are experimentally evaluated and the obtained results show the feasibility of central blood-pressure estimation and the localization of some examinees having wearable gauges with the practical accuracy and speed.}, year = {2017} }
TY - JOUR T1 - A Feasibility Study of Wearable Central Blood-Pressure Measurement System AU - Shuji Takahashi AU - Yudai Komori AU - Kyoji Nakajo AU - Kazuhiro Motegi AU - Yoichi Shiraishi AU - Hiroshi Miyashita Y1 - 2017/11/08 PY - 2017 N1 - https://doi.org/10.11648/j.jher.20170304.12 DO - 10.11648/j.jher.20170304.12 T2 - Journal of Health and Environmental Research JF - Journal of Health and Environmental Research JO - Journal of Health and Environmental Research SP - 62 EP - 71 PB - Science Publishing Group SN - 2472-3592 UR - https://doi.org/10.11648/j.jher.20170304.12 AB - A noninvasive estimation of central blood-pressure is required by the researchers or the medical practitioners in the cardiovascular treatment. In this study, first of all, a cardiovascular simulator based on the electrical circuits is described based on the model based approach and then, the representation of aorta model as a pressure transfer function is shown. This representation accelerates the simulator and then, makes it possible to estimate the central blood-pressure from a wearable gouge measurement in a real-time way. The system consisting of a main station, gateways and wearable blood-pressure gauges using one-board microcomputer is actually implemented. The functions and performances of the system are experimentally evaluated and the obtained results show the feasibility of central blood-pressure estimation and the localization of some examinees having wearable gauges with the practical accuracy and speed. VL - 3 IS - 4 ER -